LLE > Around the Lab

Past Issues

November 2002: Planar Cryo Target Positioner

A new target positioner (planar cryo) has been deployed on OMEGA to measure the properties of condensed gasses at cryogenic temperatures. It uses a closed-loop cooling system that is fully compatible with the infrastructure developed for spherical cryogenic targets. A single planar cryo system can field one target every two hours.

Planar cryo target positioner mounted on a MCTC

The target positioner is shown installed in a Moving Cryostat Transfer Cart (MCTC). (The MCTC has been partially disassembled for clarity.) For more information about the MCTC, see June 2002, Moving the MCTC.

Schematic showing how the target sits between the retaining rings of the planar cryo assembly. Illustrations, E10938

A representative target cell structure used for equation-of-state experiments. The fill gas condenses/freezes in the cell. Windows and internal components can be changed to suit the experimental requirements. Beryllium windows can be provided for transverse x-ray imaging.

Photograph of technician assembling the target cell with a pair of tweezers.

Each target cell is composed of numerous small components that must be meticulously cleaned, assembled and thoroughly tested.

Close-up view of the completed target assembly.spacerClose-up view of the top part of the planar cryo assembly. The x-ray viewing axis, fill tube, wire retainer, and retaining ring are labeled.

A close-up view shows target assembly. The target cell is clamped between a pair of retaining rings that prevent hardware from being ejected into the OMEGA chamber. Slots machined at 1 and 7 o’clock are for the fill and vent tubes. Slots machined at 4 and 10 o’clock are provided for transverse x-ray imaging.

Planar cryo target assembly prior to mounting on the assembly fixtures. The gas reservoir, split nut, target holder, and target cell are labeled.

 

A planar cryogenic target assembly is ready for filling. The coiled copper reservoir (1) is filled with room-temperature gas (e.g., D2) before attaching the target assembly to the cold head. The target holder (3) is secured to the cold head with the split nut (2), providing conductive thermal coupling.

 

Target assembly mounted on the assembly fixtures. As assembled in this image the assembly stands 12-cm high.

Assembly fixtures are used to protect targets during handling operations.

 

The target assembly at the filling station. A hand is shown in the image to show the small scale of the equipment

An assembled target (attached to an assembly fixture) is loaded into a room-temperature filling station where vacuum testing, purging and filling are completed. Final fill pressure of the reservoir is approximately one atmosphere.

The target assembly resting in the top of the target positioner.

A filled, room-temperature target is manually installed onto the target positioner. Thermal isolation between the reservoir and target cell permit the cell to be cooled below 10°K while the reservoir can exceed 250°K. The temperature differential and resulting pressure gradient ensure that the cell can be filled with the gas. Given the cell temperature, the density of the fill at shot time can be determined.

Image of sealed MCTC. The MC is approximately 3.5 feet tall when fully assembled.

After securing the target, a shroud is installed to minimize the radiative heat load incident on the target and to reduce the risk of developing condensation on the target cell. The MCTC is then sealed, evacuated, and cooled. Windows permit observation along multiple axes for positioning the target and aligning diagnostics.

Cross-sectional schematic of the target/shroud assembly. Illustrations E10960

A cross-sectional view shows the simplicity of the target/shroud assembly. Heat is conducted through the copper standoff to the cold head.

Image of a planar cryo shot in the OMEGA target chamber. The MC and several diagnostics are illuminated by the laser light hitting the target.

This photograph was taken at the instant a shot occurred. The MC is inserted into the OMEGA chamber (6 o’clock). Approximately 1.5 seconds prior to a shot, the shroud is rapidly removed from the MC. The Active Shock Breakout (ASBO) diagnostic (10 o’clock) is used to measure the velocity of laser induced shock waves that propagate through the target cell.

Close-up to the target target holder after a shot. In the center the destroyed target cell is shown.

The target cell is destroyed during the shot. The target holder and reservoir can be recovered and reused.